Differential Scanning Calorimetry (DSC)
Kindly download Material Safety Data sheet and submit the
hardcopy of filled sheet along with your sample
|
|
Make
|
: Hitachi
|
Model
|
: DSC7020
|
Specification:
|
Measurement method
|
: Heat Flux
|
Temperature Range
|
: -150 ℃ to 725 ℃
|
Scanning Rates
|
: 0.1 to 100 ℃
|
Gas Type
|
: N2, Air, Ar, O2
|
Gas Flow rates
|
: 50 to 500 ml/min
|
Cooling Units
|
: LN2 Gas cooling &
Electrical Cooling unit
|
Auto sampler
|
: 50 samples
|
|
|
 |
DSC is a thermal analytical technique that looks at how a material’s heat capacity
(Cp) is changed by temperature. It measures the amount of heat released or absorbed
by a material and hence DSC enables the measurements of the transition such as the
glass transition, melting, and crystallization.
Heat Flux DSC A technique in which the sample and reference crucible are
heated uniformly (under the same conditions) with a definite amount of heat. The
difference in temperature between the sample and the reference material gives the
heat capacity of the sample and is measured as a function of time and temperature.
|
Glass Transition temperature
Melting and Crystallization temperature
Percent Crystallinities
Compositional analysis of blends
Thermal histories
Thermal denaturation for proteins
Purity Determination
Polymorphism
Drug Compatibility studies, etc.
|
Contact : 022-2159 6866/89 Email Id :
nmr600@iitb.ac.in
|
|
|
Charges (including 18% GST)
*For Thermal Analysis, standard ramp is 10 ℃/min. Heating rates
less than standard rate will be charged by hours.
|
|
Academic |
National Lab |
Industry |
IIT Bombay |
Heating Program: Temp. Range @ 10
℃/min |
Rs. |
Rs. |
Rs. |
Rs. |
|
|
|
|
RT- <400 ℃ |
826
|
2478
|
4130
|
350
|
RT- <600 ℃ |
944
|
2832
|
4720
|
400
|
RT- 700 ℃
|
1062
|
3186
|
5310
|
450
|
Low Temp. -80 to 25 ℃ (RT)
|
944
|
2832
|
4720
|
400
|
Low Temp. -150 to 25 ℃ (RT)
|
1888
|
5664
|
9440
|
800
|
Cp Measurement (RT- 400 @ 4 ℃/min)
|
4248
|
10,620
|
21,240
|
1800
|
|
|