Liquid Nitrogen Plant

  Instrument Details  

Make : Stirling Cryogenic Pvt. Ltd, Netherlands.
Model : StirLIN-1 Compact
Model 1 ~10LPH capacity.
Specification :
LN2 Plant – 10LPH (StirLIN-1 compact)
Chiller Cooling capacity - 12 kW
Refrigerant - R407C
Suction/Discharge pressure - 3.7/16.5 bar
Suction temperature - 12℃
Cryogenerator - power 45 kW
Production capacity - 10 lph
Refrigeration capacity - 4.4 kW@65K
Working gas Helium Filling pressure - 22 barg@50Hz
Liquid nitrogen purity - >99%
Air compressor type - Screw compressor
Motor rated power – 5.5 kW
Working pressure - 7.5 bar





  Working Principles   

         A PSA technique is used to extract nitrogen from compressed ambient air, which contains 78% nitrogen. The nitrogen is then condensed and kept in a storage container under positive pressure as a liquid.



  Faculty Incharge   

         Prof. Suparna Mukherjee, Head SAIF


  Applications   

        The LN2 most commonly used for cryopreservation of biological material.
        It is used to precool liquid helium in NMR to minimize leaks of L He. Also, it
        is used in High temperature superconductors etc.