Liquid Chromatograph Mass Spectrometer
Kindly download Material Safety Data sheet and submit the hardcopy of filled sheet along with your sample
|
|
| Make |
: Varian Inc, USA |
| Model |
: 410 Prostar Binary LC with 500 MS IT PDA Detectors.
|
| Specification |
: 1) Direct Infusion Mass with ESI & APCI Negative & Positive mode ionization, mass ranging from 50 to 2000 m/e
2)LCMS / MS & MS^n ION TRAP
3)HPLC with PDA Detector
4)HPLC PDA Detector - Massspectrometer |
|
 |
|
Liquid chromatography / Mass Spectroscopy (LC / MS) is a technique which combines high performance liquid chromatography HPLC, a powerful analytical separation technique with mass spectroscopy ,a powerful analysis & detection technique. There are two common atmospheric pressure ionization (API) LC/MS process: Electrospray Ionization (ESI) & Atmospheric Pressure Chemical Ionization (APCI). Both are soft ionization technique. Both of these processes are compatible with most chromatographic separations. API processes are also compatible with most mobile phase solvents (volatile) & volatile buffers.
HPLC can be combined with Photo Diode Array detector (PDA).
HPLC seperations can be detected by PDA & Mass spectrometer as different detectors.
Ion Trap MS/MS (MS n) Function :
A. Ion Injection & accumulation
B. Isolation( Ejection of masses above & below parent ion)
C. Fragmentation( CID of precursor of all product ions)
D. Ejection of product ions to detector
Ion trap facility has it's own advantage over other tandem mass as in this method you can select your interested ion peak and accumulate for fragmentation for further MS study so this will be without any interference from other ions results in higher sensitivity. This advantage is exploited for small impurities profiling studies.
PDA Detector :
It has three dimensional projection over entire UV-VIS range.
It is non-distructive detector so it can be put in series like HPLC-PDA Mass Spectrometer.
|
|
|
MAIN FEATURES OF THE INSTRUMENT :
|
Mass Range : 50-2000 amu
Binary HP- LC system with PDA detector.
Ionisation Method : API 1)ESI positive & negative 2)APCI positive & negative
Direct Infusion for Mass Analysis ( MS , MS /MS, MS^n )
Ion Trap facility to study selected mass over other interference & fragmentation pattern
Binary HP- LC system with Mass as detector.
Binary HP- LC system with PDA detector.
HPLC PDA Mass Spectrometer.
|
Qualitative Analysis
Quantitative Analysis
Impurity Profiling
Metabolite Studies
Pharmacokinetics
Pharma & biomedical applications:
|
|
1. Expected structure or mass can be provided.
2. Solvent for mobile phase should be mentioned or solubility of compound in volatile solvent should be given.
3. As HP-LC is available in most departments, LC of submitted samples indicating the operating conditions should be supplied along with the sample.
4. At present C8,C18 & others like cyano,amino,Si columns are available. You can submit the column of your interest for separation or repeatability of HP-LC run.
5. For HPLC PDA run frequency should be given.
6. Source of the compound should be mentioned.
7. MS-DS (Material Safety Data Sheet) should be given along with samples to ensure that there are no toxic sample being given. Samples should not be toxic or hazardous. Samples will not be accepted unless accompanied by MS-DS.
|
Contact : 022-21596873 Email Id : pradnyaj@iitb.ac.in
|
|
Liquid Chromatograph Mass Spectrometer Charges includes GST
|
| | Industry | University | National Lab/R&D's | |
| HPLC with PDA detector | 7080/- | 1328/- | 3186/- | Per sample
|
HPLC + PDA detector + Mass Spectrometer : For this analysis please refer to HRLCMS-QTOF
|
LCMS with APCI Or ESI : For this analysis please refer to HRLCMS-QTOF
|
|
|